Menu

Genie U, Ultrapure & RO Lab water system, RephiLe

Two-in-one integrated system. Quality of ultrapure water meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.
The system is manufactured in an ISO 9001 : 2015 certified manufacturing site.

Clear

Description

<strong>Product No</strong><strong>Specification</strong><strong>Download</strong><strong>Application</strong><strong>Features</strong>

Product No. Description
RG0U01000K Genie U 12 System Set
RG0U010T0K Genie U 12 System Set, with TOC
RG0U02000K Genie U 24 System Set
RG0U020T0K Genie U 24 System Set, with TOC
RG0U03000K Genie U 32 System Set
RG0U030T0K Genie U 32 System Set, with TOC
Set Included System includes:

  1. Main system
  2. Control Console
  3. RO membranes
  4. UV preinstalled

 Feed Water Requirement  
 Feed Water Conductivity / TDS  <2000 μS/cm / <1000ppm
 Operating temperature   5~45 ℃
 Feed Water Pressure   1 – 6 kg/cm2  (15 – 90 psi)
 Product Water Quality  
 RO Water  
 Flow Rate  12, 24, 32 L/hr
 Dispenser Rate  0 – 2 L/min
 Conductivity (@25°C)  Typically < 20 μS/cm (feed water < 2000 μS/cm)
 Ultrapure Water
 Dispenser rate   0 – 2 L/min
 Resistivity (@25℃ )   18.2 MΩ·cm
 TOC*   < 5 ppb
 Particles ( > 0.2 μm)**   No Particles with size > 0.22 μm
 Microorganisms***  < 0.1 cfu/ml
 Pyrogens (endotoxins)****  < 0.001 Eu/ml
 RNAse****  < 0.5 pg/ml
 DNAse****  < 10 pg/ml
 Dimensions  
 Main System: Length x Depth x Height 32cm x 44cm x 54cm
 Main System: Length x Depth x Height 21cm x 29cm x 61cm

* Product water quality may vary due to local feed water conditions.
** with a 0.2 μm final filter
*** with a 0.2 μm final filter or RephiBio filter
**** with a RephiBio filter

With Ultrapure Water

• HPLC mobile phase preparation

• Preparation of reagent blank solutions

• As sample diluent for GC, HPLC, ICP-MS, AA and other analytical techniques

• Preparation of buffers and culture media for mammalian cell culture

• Preparation of molecular biology reagents, etc.

With RO Water

• Glassware cleaning

• Washing machine for glassware

• Water bath water

• Autoclave

• Feed water for laboratory animals

Wireless communication amongst components providing unlimited possibilities

RFID tracking of consumables and RO membranes to ensure optimal system performance

On-line TOC measurement based on complete oxidation methodology

Stable RO permeability over a wide range of water temperatures

A full range of cartridges for various applications including ultra-low TOC, low Mg, low boron, ICP and DI type

Choices of final filters to remove specific contaminants

Tank water level displayed from the continuous liquid level sensor of the storage tank

Automatic system shut-off upon detection of any water leakage

Optional tank circulation mode to keep the water quality stable in the tank