BEING Forced Air Ovens have an integrated fan that actively moves the air inside the chamber – resulting in an even temperature
distribution throughout the chamber. One benefit is optimal temperature uniformity for reproducible results, for example in material testing and for drying protocols with very tight temperature requirements. Another advantage is a much faster drying process compared to natural convection. When opening the door, the temperature will recover faster to the set temperature level in a mechanical convection oven.
BO-F Series – Touch Screen | BO-FL Series – LCD
Model | BO-30F BO-30FL |
BO-50F BO-50FL |
BO-120F BO-120FL |
BO-200F BO-200FL |
BO-400F BO-400FL |
Volume | 35L | 59L | 115L | 234L | 400L |
Temperature Range | Amb+10°C ~200°C (Max 300°C) | Amb+10°C ~200°C (Max 300°C) | Amb+10°C ~200°C (Max 300°C) | Amb+10°C ~200°C (Max 300°C) | Amb+10°C ~200°C (Max 300°C) |
Temperature resolution | 0.1°C | 0.1°C | 0.1°C | 0.1°C | 0.1°C |
Uniformity | ±2.0°C | ±2.0°C | ±2.0°C | ±2.0°C | ±2.0°C |
Heating Time (Amb~100°C) | 25mins | 25mins | 25mins | 25mins | 25mins |
Standard/Max Number of shelvings | 2 (5) | 2 (6) | 2 (10) | 2 (16) | 3 (16) |
Max Load per shelf | 20KG | 20KG | 20KG | 20KG | 20KG |
Net Weight | 43KG | 51KG | 83KG | 112KG | 210KG |
Timer Range | 1~5999 mins | 1~5999 mins | 1~5999 mins | 1~5999 mins | 1~5999 mins |
Internal Dimension WxDxH(mm) | 320x300x320 | 400x305x415 | 520x430x530 | 650x500x650 | 1000x510x800 |
External Dimension WxDxH (mm) | 610x550x540 | 690x560x640 | 810x685x755 | 940x750x875 | 1285x750x1060 |
Power | AC220V,50Hz 900W |
AC220V,50Hz 1100W |
AC220V,50Hz 2050W |
AC220V,50Hz 2500W |
AC220V,50Hz 3100W |
Note: Measured by H2O, the viscosity of the liquid will vary from stirring capacity.
Humanized design
• With reservation and timing functions, there is no need to wait, which effectively improves the efficiency of the experiment.
• Available in Chinese and English menus to meet different language needs, with °C / °F conversion.
• The height of the shelf is adjustable to meet different cultivation needs.
• Professional stacking foot design, the machine can be stacked, saving lab space and improving the use efficiency (same specification).
Safety
• Protection on instruments: Comply international standard secondary temp limiting alarm system, When the heating is out of control or exceeds the maximum limit temperature, the heating is automatically cut off with sound and light alarm, ensure operating is safe without any accident.
• Protection on key components: Key components are equipped with over-current, over-temperature, overload and other safety protection to prevent accident.
• Protection on samples: When the temperature inside the box is higher or lower than the set temperature, the alarm will start to cut off the heater, with sound and light alarm to remind the operator to protect the sample from normal test without accident.
• Protection on operation: The cabinet and door are specially insulated to make the surface temperature of the cabinet low, ensuring the operator’s safety and no accidents.
• Breakdown message provided: When the instrument breakdown, the message will shown on the screen to help operator clearly check.
Convenient data processing
• The touch screen type standard with a USB interface, which can record the change status of the temperature parameters.
• The LCD screen is option for USB interface.
• It can be equipped with RS232 data interface, which can realize remote control of the machine through software (option).
• RS-232 and USB can be selected one of them.