Genie PURIST ultrapure water system produces Type I ultrapure water from RO, distilled or deionized water. The system is manufactured in an ISO 9001 : 2015 certified manufacturing site.
* It’s recommended to use a pressure regulator if feed water pressure is above 15 psi (1 bar) • HPLC mobile phase preparation • Preparation of reagent blank solutions • As sample diluent for GC, HPLC, ICP-MS, AA and other analytical techniques • Preparation of buffers and culture media for mammalian cell culture • Preparation of molecular biology reagents, etc. Wireless communication amongst components providing unlimited possibilities
Product No.
Description
RG0S00000K
Genie PURIST System Set
RG0S000T0K
Genie PURIST System Set, with TOC
Set Included
System includes:
Feed Water Requirement
Feed Water
EDI Water, RO Water, Distilled Water, Deionized Water
Operating temperature
5~45 ℃
Feed Water Pressure
0 – 1 kg/cm2 (0 – 15 psi)
Product Water Quality
Flow Rate
0 – 2 L/min
Resistivity (@25℃ )
18.2 MΩ·cm
TOC Level**
< 5 ppb
Particles ( > 0.2 μm)***
No Particles with size > 0.22 μm
Microorganisms****
0.1 cfu/ml
Pyrogens (endotoxins)*****
< 0.001 Eu/ml
RNases*****
< 0.5 pg/ml
DNases*****
< 10 pg/ml
Dimensions
Main System: Length x Depth x Height
30cm x 48cm x 51cm
Dispenser: Length x Depth x Height
21cm x 29cm x 61cm
** When TOC Level of feed water is < 50 ppb
*** with a 0.2 μm final filter
**** with a 0.2 μm final filter or RephiBio filter
***** with a RephiBio filterWith Ultrapure Water
RFID tracking of consumables and accessories to ensure optimal system performance
On-line TOC based on complete oxidation methodology
A full range of cartridges for various applications including ultra-low TOC, low Mg, low boron, ICP and DI type
Eliminating bacteria and reducing TOC to trace level by a dual wavelength UV lamp
Choices of final filters to remove specific contaminants
A universal water level sensor to prevent the system from pumping air
An optional automatic system shut-off upon detection of any water leakage